Lifeguard AED equipment is more than the device. At an aquatic facility the accessories are what make it usable, because your casualty arrives wet.
Status indicator green. Pads sealed, in date, and a spare set present. Battery in date. Towel and razor in the case. Case unobstructed and signed. This takes fifteen seconds and it is the difference between equipment and decoration.
Within reach of the water, unlocked and signposted. An AED in a locked office is not part of the response. Larger facilities place one per venue rather than one per site, because the time to fetch it across a property is time compressions are not being supported.
Replace the pads immediately, even if they were applied and no shock was delivered. Pads are single use. Retrieve the event data if the unit stores it, because the receiving hospital will want it.
The conductive gel dries out, and dry pads do not conduct. This is the failure that looks invisible: the device passes its self-test and the pads are useless. Checking dates is the only way to catch it.
The AED unit, adult pads with a spare set, paediatric pads or mode, a towel for drying a wet chest, a razor, gloves and a barrier mask, and a spare or in-date battery.
Because pads will not adhere to a wet chest, and a film of water across the chest can carry the shock over the skin rather than through the heart.
The conductive gel dries out and dry pads do not conduct. The device still passes its self-test, so checking the pad dates is the only way to catch it.
Reviewed by Jackie Stern, Lifeguard Instructor, Firefighter and EMT. Jackie Stern has taught lifeguarding for more than 20 years and works as a firefighter, EMT and swim instructor. She reviews course content, prerequisites and rescue standards against current certification requirements.